IMR Press / FBL / Volume 11 / Issue 3 / DOI: 10.2741/2008

Frontiers in Bioscience-Landmark (FBL) is published by IMR Press from Volume 26 Issue 5 (2021). Previous articles were published by another publisher on a subscription basis, and they are hosted by IMR Press on imrpress.com as a courtesy and upon agreement with Frontiers in Bioscience.

Article

Simultaneous In Vivo Measurement of Dopamine, Serotonin and Ascorbate in the Striatum of Experimental Rats Using Voltammetric Microprobe 

Show Less
1 Analytical Chemistry Laboratory, National Institute of Chemistry, Hajdrihova 19, SI-1001, Ljubljana, Slovenia
2 Institute of Pathophysiology, Faculty of Medicine, University of Ljubljana, Zaloška 4, SI-1000, Ljubljana, Slovenia
3 Institute of Histology and Embryology, Faculty of Medicine, University of Ljubljana, Korytkova 2, SI-1000, Ljubljana, Slovenia
4 Eye Clinic, University Medical Center, Zaloška 29, SI-1000, Ljubljana, Slovenia
Front. Biosci. (Landmark Ed) 2006, 11(3), 2782–2789; https://doi.org/10.2741/2008
Published: 1 September 2006
Abstract

The in vivo performance of a voltammetric microprobe based on overoxidized poly(1,2-phenylenediamine) coated carbon fiber microelectrode (OPPD/CFME), developed in our laboratory, is presented. For this purpose, an OPPD microprobe was stereotaxicaly implanted in the striatum of a deeply anesthetized Wistar rat for the simultaneous measurement of dopamine, serotonin and ascorbate. Furthermore, the post mortem levels of these physiologically important compounds were monitored after the rats were terminated with an overdose of anesthetic introduced through an indwelling jugular catheter. Using cyclic (CV) and square-wave (SWV) voltammetry, the OPPD/CFME was demonstrated to exhibit efficient separation of the voltammetric signals of dopamine, serotonin and ascorbate in the presence of biological matrix, with the SWV mode allowing more convenient detection regarding both sensitivity and selectivity. Explicit proof of in vivo dopamine detection at the OPPD/CFME was achieved via the absence of the dopamine signal in rats with unilateral lesions of nigrostriatal dopaminergic neurons, which was induced by the use of the selective dopaminergic neurotoxin 6-hydroxydopamine (6-OHDA). In stark contrast to non-treated rats, where a strong signal corresponding to 1.2 µM dopamine was measured at ca. 90 s after the rats' death, the signal for dopamine in dopamine-depleted striatum of 6-OHDA rats was absent. A critical comparison of the in vivo performance of the OPPD/CFME to that of bare and Nafion-coated carbon fiber microelectrodes, often used in neurophysiological studies, clearly showed a significant advantage of the former microprobe. The OPPD/CFME allowed multiple and repetitive in vivo measurements, along with pre- and post-measurement external calibration, with no loss in selectivity and an acceptable loss in sensitivity, indicating that the active sensing sites were not adversely blocked by the components of the extracellular fluid, thus affirming the great practical in vivo applicability of the OPPD microprobe.

Keywords
Overoxidized Poly(1
2-Phenylenediamine)
Carbon Fiber
Microelectrode
Dopamine
Serotonin
Ascorbate
In Vivo
Brain Voltammetry
Striatum