IMR Press / RCM / Volume 24 / Issue 7 / DOI: 10.31083/j.rcm2407191
Open Access Systematic Review
Evaluating the Efficacy and Safety of the Thumper Device for Cardiac Arrest: A Systematic Literature Review and Meta-Analysis
Show Less
1 International Nursing School of Hainan Medical University, 571199 Haikou, Hainan, China
2 Key Laboratory of Emergency and Trauma, Ministry of Education, 571199 Haikou, Hainan, China
3 Research Unit of Island Emergency Medicine, Chinese Academy of Medical Sciences, 571199 Haikou, Hainan, China
4 The Second Affiliated Hospital of Hainan Medical University, 571199 Haikou, Hainan, China
5 Department of Physical Education Hainan Medical University, 571199 Haikou, Hainan, China
6 Faculty of Physical Education, Yan’an University, 716000 Yan’an, Shaanxi, China
*Correspondence: zhanghuashelley@hotmail.com (Hua Zhang); 1490560829@qq.com (Jiameng Wang)
Rev. Cardiovasc. Med. 2023, 24(7), 191; https://doi.org/10.31083/j.rcm2407191
Submitted: 21 October 2022 | Revised: 16 December 2022 | Accepted: 16 January 2023 | Published: 3 July 2023
Copyright: © 2023 The Author(s). Published by IMR Press.
This is an open access article under the CC BY 4.0 license.
Abstract

Background: Cardiopulmonary resuscitation (CPR) is a major rescue measure for cardiac arrest (CA) patients, and chest compression is the key to CPR. The Thumper device was designed to facilitate manual compression during CPR. However, current randomized controlled trials (RCTs) provide controversial findings on the efficacy of the Thumper device. Objectives: This meta-analysis aimed to compare the clinical benefits of using the Thumper device with manual chest compressions during the provision of CPR for patients in CA. Methods: Relevant studies were retrieved from various databases, including Ovid, PubMed, Web of Science, EMBASE, Cochrane, and CNKI, and by manually searching the reference lists of research and review articles. All RCTs published in either English or Chinese until June 31, 2020, were included in the meta-analysis. The odds ratios (ORs) and their 95% confidence intervals (95% CIs) for the return of spontaneous circulation (ROSC), survival rate (SR), and the incidence of rib fractures (RFs) were compared between the manual and Thumper chest compressions. Results: A total of 2164 records were identified, of which 16 were RCTs with an overall risk of bias ranging from low to medium classification. Following CPR, the odds ratios for ROSC, SR, and RF were significantly better for the Thumper chest compression with ORs of 2.56 (95% CI 2.11–3.11, I2 = 0%), 4.06 (95% CI 2.77–5.93, I2 = 0%), and 0.24 (95% CI 0.14–0.41, I2 = 0%), respectively. Conclusions: The Thumper compression devices may improve patient outcome, when used at inhospital cardiac arrest. This review suggests a potential role for mechanical chest compression devices for in-hospital cardiac arrest, but there is an urgent need for high-quality research, particularly adequately powered randomised trials, to further examine this role.

Keywords
cardiac arrest
cardiopulmonary resuscitation (CPR)
thumper
manual chest compression
meta-analysis
emergency care
1. Introduction

Cardiac arrest (CA) is a medical emergency caused by the abrupt loss of heart function resulting in a sudden loss of blood flow [1]. CA is the leading cause of death worldwide. In the United States and Europe, more than 300,000 and 450,000 people, respectively, die of CA each year [2, 3]. The performance of timely cardiopulmonary resuscitation (CPR) determines the survival rate as well as the neurological outcome in patients suffering from CA [4, 5]. The quality of the chest compression is essential during CPR to maintain organ perfusion which ultimately determines the prognosis of CA patients.

Chest compressions can be provided manually or mechanically. The main advantage of manual chest compression is that it can be administered immediately at the scene and, hence, improves the chances of survival [5]. However, the CPR procedure is also physically demanding. During prolonged CPR, the rescuer’s fatigue can reduce the quality of the chest compressions, particularly if the procedure is performed outside the hospital or on hospital transport [6].

In order to overcome the limitations of manual chest compressions, several mechanical compression devices have been proposed. Namely point chest compression, suction chest compression, full chest coverage load distribution compression (vest type), and broad chest compression (three-dimensional type) [7]. Some experimental studies have shown that mechanical chest compression can provide more uniform recoil and increase the intrathoracic pressure when compared with manual chest compression, thus increasing the effective coronary perfusion pressure and systemic blood flow [7, 8, 9]. This provides an advantage when the CPR is performed outside the hospital or on ambulances [10]. The Thumper compression device uses point chest compression and is seeing an increase in use in the rescue of CA patients. When using a Thumper device, the rescuers can focus on providing supplementary advanced life support, as they no longer need to worry about providing manual chest compressions. This may improve the efficiency and outcomes of the rescue operation. Additionally, compared with manual compression, the Thumper device provides constant high-quality chest compression, frequency, depth and rhythm. This eliminates the inconsistencies caused by operator fatigue following a prolonged manual resuscitation and changing operators throughout the procedure.

However, the clinical benefit of using the Thumper device as opposed to the manual compression is still not clear. There is an obvious inconsistency in the published literature on the efficacy of Thumper compression. Various randomized controlled trials (RCTs) can not demonstrate that Thumper compression can improve survival when compared with manual compression. Nevertheless, several observational studies showed that Thumper device compression could improve the survival rate [11, 12]. In addition, there is a lack of data on the safety profile of the Thumper device. Therefore there is a need for a meta-analysis to summarize the relative effectiveness and safety of Thumper devices in relation to manual compression in patients with CA. In view of this, we conducted a meta-analysis to comprehensively evaluate the effects of Thumper compression and manual compression in patients with cardiac arrest.

2. Materials and Methods

The meta-analysis was conducted in accordance with the agreement registered in the PROSPERO database on September 25, 2020 (Registration Number: CRD420206025).

2.1 Information Sources

Relevant studies published before June 30, 2020, were retrieved from the Ovid, PubMed, Web of Science, EMBASE, Cochrane Library, Trials registries, Google Scholar, and the China National Knowledge Infrastructure (CNKI) electronic databases. In addition, snowballing was used to identify relevant research articles from the reference list of published studies. The Google Academy was used to identify and screen studies that cited such evidence. On April 28, 2020, we conducted a search on Google Academy and conducted a supplementary search on the websites of relevant organizations, including government departments and research institutions. The database search was again updated on November 15, 2022, and a final literature and snowball search was performed on November 22, 2022.

2.2 Search Strategy

Twelve known related studies [1, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21] were used to identify records in the electronic databases. We determined the candidate search terms by screening the records’ titles, abstracts, and search processes. These terms were then used to formulate a draft of the search strategy, and other search terms were determined according to the results of the strategy. The PubReMiner tool was used to identify and check the frequency of the search terms. The MEDLINE policy uses the Cochrane RCT filter reported in the Cochrane manual version 5.2. The search strategy was limited to only English or Chinese language articles and there was no restriction on publication. The search strategy was then verified by ensuring that it could identify the 12 known related studies on the PubMed and EMBASE databases. The search strategy was developed by experienced researchers among the project members. The overall structure and accuracy of the final search strategy were discussed and peer-reviewed by Zhang. The final search strategy used a combination of keywords to describe the condition (cardiac arrest), intervention (compression device), and study design (RCT).

2.3 Inclusion and Exclusion Criteria

All RCTs on adult (age 18 years) CA patients published in either English or Chinese that compared clinical outcomes between the Thumper and manual chest compressions were included in the meta-analysis. The primary clinical outcome measure for this meta-analysis was the return of spontaneous circulation (ROSC), and the secondary outcome measures were discharge survival rate and the incidence of rib fractures. If these all of the outcomes indicators were not published in the study, the corresponding author of the research articles was contacted via e-mail to provide the additional data. If these data were not provided, the article was excluded.

All studies using mechanical chest compression with mechanical devices other than the Thumper were excluded. Additionally, studies that included children under 18 years, animal and simulation studies, non-RCTs, and those lacking controls were excluded. Unpublished manuscripts and conference abstracts were also excluded.

2.4 Selection Process

Two researchers (Luo and Zhang) independently reviewed the titles and abstracts of all records. The researchers then filtered the titles and abstracts of all retrieved articles to identify the full-text articles for inclusion. If the included articles met the eligibility criteria, the full text was further searched to ensure that all required data were available. The references of the articles were also reviewed to identify additional suitable studies. Any disagreements in the retrieved articles were resolved through discussion. If the researchers failed to reach an agreement, a third researcher (Wang) was consulted for the final decision.

2.5 Data Collection Process

The general information of the research article (first author name, year of study, and publication date), the sample size of the control group and experimental group, and the type of study were extracted from the articles. Additionally, the American Heart Association (AHA) guidelines used to define the control group, the intervention measures of the experimental and control groups, and the outcome indicators ROSC and discharge survival and the incidence of rib fractures were also extracted. All the data were extracted by three researchers (Luo, Zhang, and Wang) and recorded on an excel sheet. The extracted data were compared, and any differences were resolved through discussion. Finally, one of the researchers (Luo) entered the data into the Review Manager (ReMan) 5.3 software (Nordic Cochrane Center, London, UK) and checked their accuracy. If any part of the above data were not clearly described in the research article, the corresponding author was consulted to provide further details.

2.6 Risk of Bias Assessment

The included studies were assessed for bias in accordance with the evaluation of the authenticity of RCT criteria published by the Cochrane Collaboration Network. This assessment is based on seven criteria: random sequence generation, allocation concealment, blinding of participants and personnel, blinding of outcome assessment, incomplete outcome data, selective reporting, and other biases. Two reviewers (Luo and Zhang) independently applied the risk assessment tools to all included studies and classified the risk bias of each study as low-risk, high-risk, or unclear. A justification for the classification was also provided. Any differences in determining the risk of bias or justification were resolved through discussions between the two reviewers (Luo and Zhang). If necessary, a third reviewer (Wang) acted as an arbitrator.

Funnel plot analysis was used to estimate the publication bias. If the funnel plot was asymmetric, the research articles were reviewed to identify possible sources of bias, such as publication or trial design bias. A sensitivity analysis was performed to determine the robustness of the observed outcomes.

2.7 Statistical Analysis

The data were processed using the Revman 5.3 developed by the Cochrane Collaboration. The odds ratios (OR) or mean differences (MDs) were reported for dichotomous and continuous variables. The I-squared (I2) test was used to assess the heterogeneity of the included studies. An I2 greater or equal to 50% indicates high heterogeneity between studies [22]. If the statistical heterogeneity among studies was high, the subgroup analysis was deemed invalid, and the random effect model (RE) was used to analyze the statistical indicators. Conversely, the fixed effect (FE) model was used to analyze the statistical indicators if the statistical heterogeneity was low. The differences in clinical outcomes between the two compression methods were deemed statistically significant if the p-value was below 0.05.

2.8 Ethical Considerations

This study was conducted in compliance with the recommendations published by the preferred reporting item of the guidelines (PRISMA 2020) for meta-analysis and systematic reviews. Since no patient data were collected in this study, ethical approval was not required.

3. Results
3.1 Searching Results/Study Selection

The studies identified during the literature screening process are summarized in Fig. 1. A total of 2164 records were retrieved in our database search, of which 11 articles were obtained by snowballing. After removing the duplicate items, we identified 1454 relevant records, of which only 45 were full-text articles. From these full-text articles, only 16 papers met the eligibility criteria stated above. Later, we searched for all references that ultimately included evidence. However, no other articles identified through this search met the eligibility criteria.

Fig. 1.

2020 PRISMA flowchart of the literature search and study selection process for new systematic reviews, which included searches of databases, registers, and other sources. RCT, randomized controlled trials.

3.2 Study Characteristics and Quality

This meta-analysis included 16 RCTs involving 2275 CA patients. The characteristics of these studies and the participants are summarized in Table 1 (Ref. [23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37, 38]). All 16 articles used the random grouping method, and 4 of them [23, 24, 25, 26] reported the exact random method used. The Cochrane bias risk assessment tool was used for quality assessment and bias risk assessment was done at study level, all 16 articles had an overall risk of bias, ranging from low to high, as shown in Table 2 (Ref. [23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37, 38]) and Fig. 2.

Table 1.Characteristics of the included trials and participants.
Author and year Country Setting Year of study Sample (T/C) Study design Age (T/C) Men (%) AHA version ROSC (T/C) Discharge Survival rate (T/C) Rib fracture (T/C)
Meng LF 2019 [23] China ER 2015–2018 200 (100/100) RCT (41.03 ± 3.11)/ (42.01 ± 2.17) 56.5 2005 76/46 59/24 NR
Zheng H 2019 [27] China ER 2018–2018 100 (48/52) RCT 61 61 2015, 2010 27/18 NR NR
Zhang CY 2017 [28] China ER 2015–2017 150 (70/80) RCT (59.18 ± 11.98)/ (62.65 ± 8.57) 78.7 2010 48/41 NR NR
You Y 2017 [29] China ER 2015–2016 80 (40/40) RCT (63.71 ± 13.97)/ (67.74 ± 15.21) 60 2015 32/20 NR 0/4
He NN 2016 [24] China IHCA 2005–2011 400 (200/200) RCT 43.2 ± 5.33 72 2005, 2010 166/131 NR NR
Gong N 2016 [25] China ER 2010–2014 247 (112/135) RCT (61 ± 17)/ (61 ± 18) 62.3 2010 Not applicable 18/4 NR
Dong QL 2016 [30] China IHCA 2012–2015 100 (50/50) RCT (64.12 ± 5.07)/ (63.71 ± 4.23) 56 2005 46/28 NR 0/6
Liu HL 2015 [31] China IHCA 2012–2014 135 (65/70) RCT (68.37 ± 11.16)/ (67.99 ± 11.09) 59.2 2010 42/35 NR 2/9
Ding HB 2014 [26] China ER 2011–2014 68 (34/34) RCT (45 65)/ (46 68) 54.4 2005 23/13 16/6 0/5
Guo S 2014 [32] China IHCA 2009–2012 158 (80/78) RCT (51.23 ± 9.86)/ (50.30 ± 10.18) 58.9 2010 43/24 NR 1/4
Jin Y 2013 [33] China IHCA 2008–2012 146 (71/76) RCT 14 75 47.8 2008 31/21 NR NR
Liu JF 2013 [34] China IHCA 2013 32 (16/16) RCT 39 75 59.3 2010 5/3 NR NR
Hu PB 2012 [35] China ER 2010–2012 107 (55/52) RCT (47.38 ± 13.35)/ (45.44 ± 12.43) 52.3 2005 30/18 NR 7/20
Huang Q 2011 [36] China IHCA 2008–2009 152 (62/91) RCT 19 76 63.1 2005 25/12 10/4 NR
Lu XG 2010 [37] China ER 2009–2010 150 (74/76) RCT (47.72 ± 14.25)/ (45.50 ± 13.82) 59.3 2005 42/28 25/11 2/8
Taylor 1978 [38] America ER —– 50 (26/24) RCT 10/10 3/2 7/8

NR, not reported; RCT, randomized controlled trial; ROSC, return of spontaneous circulation; T/C, Thumper group/Control group; ER, emergency room; IHCA, in-hospital cardiac arrest; AHA, American Heart Association.

Table 2.Evaluation of risk assessment of the included RCT according to the Cochrane Collaboration Network.
Author and year 1. Random sequence generation 2. Allocation concealment 3. Blinding of participants and personnel 4. Blinding of outcome assessment 5. Incomplete outcome data 6. Selective reporting 7. Other bias
Meng LF 2019 [23] Low risk Unclear risk Low risk unclear Low risk Low risk High risk
Zheng H 2019 [27] Low risk Unclear risk High risk unclear Low risk Low risk High risk
Zhang CY 2017 [28] Low risk Unclear risk Low risk Low risk Low risk Low risk High risk
You Y 2017 [29] Low risk Unclear risk High risk unclear Low risk Low risk High risk
He NN 2016 [24] Low risk Unclear risk Low risk unclear Low risk Low risk High risk
Gong N 2016 [25] Low risk Unclear risk Low risk Low risk Low risk Low risk High risk
Dong QL 2016 [30] Low risk Unclear risk Low risk unclear Low risk Low risk High risk
Liu HL 2015 [31] Low risk Unclear risk Low risk unclear Low risk Low risk High risk
Ding HB 2014 [26] Low risk Unclear risk Low risk Low risk Low risk Low risk High risk
Guo S 2014 [32] Low risk Unclear risk High risk unclear Low risk High risk High risk
Jin Y 2013 [33] Low risk Unclear risk High risk High risk Low risk Low risk High risk
Liu JF 2013 [34] Low risk Unclear risk Low risk Low risk Low risk Low risk High risk
Hu PB 2012 [35] Low risk Unclear risk Low risk High risk Low risk Low risk High risk
Huang Q 2011 [36] Low risk Unclear risk Low risk unclear Low risk Low risk High risk
Lu XG 2010 [37] Low risk Unclear risk Low risk High risk Low risk Low risk High risk
Taylor 1978 [38] Low risk Unclear risk High risk unclear High risk unclear High risk

RCT, randomized controlled trial.

Fig. 2.

Risk of bias summary for included studies.

3.3 ROSC Rate

Among the 16 RCT studies, 15 studies [23, 24, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37, 38] compared the ROSC between the mechanical and manual chest compressions. Thirteen of these studies [23, 24, 26, 27, 28, 29, 30, 31, 32, 33, 35, 36, 37] showed that ROSC in the Thumper chest compression group was significantly higher when compared with the manual chest compression group (p < 0.05). In contrast, no significant difference was reported in the other two studies conducted by Taylor et al. [38] and Liu et al. [34]. The meta-analysis showed that the ROSC in the Thumper chest compression group was better than that in the manual chest compression group (OR = 2.56, 95% CI 2.11~3.11, Z = 9.47, p < 0.05). The forest plots of this analysis are presented in Fig. 3.

Fig. 3.

The ROSC forest plot of Thumper chest compression versus manual chest compression. CPR, cardiopulmonary resuscitation; CI, confidence interval; ROSC, return of spontaneous circulation.

3.4 Discharge Survival Rate

Among the 16 RCT studies, six studies compared [23, 25, 26, 36, 37, 38] the discharge survival rate. Five of these studies [23, 25, 26, 36, 37] showed that the discharge survival rate in the Thumper chest compression group was significantly higher (p < 0.05) than that of the manual chest compression group, while no significant difference (p > 0.05) was noted in the study by Taylor et al. [38]. The meta-analysis results showed that the discharge survival rate in the Thumper chest compression group was significantly higher when compared with the manual chest compression group (OR = 4.06, 95% CI 2.77–5.93, Z = 7.22, p < 0.05). The forest plot analysis is presented in Fig. 4.

Fig. 4.

The discharge survival rate forest plot of Thumper chest compression versus manual chest compression. CPR, cardiopulmonary resuscitation; CI, confidence interval.

3.5 Incidence of Rib Fractures

Among the 16 RCTs included, eight studies [26, 29, 30, 31, 32, 35, 37, 38] compared the incidence of rib fractures. Seven of these studies [26, 29, 30, 31, 32, 35, 37] showed that the incidence of rib fractures in the Thumper chest compression group was significantly lower (p < 0.05) when compared with the manual chest compression group. There was no heterogeneity among the studies (p = 0.67, I2 = 0%, I2 <50%), so the FE model was used for meta-analysis. The meta-analysis results showed that the incidence of rib fractures in the Thumper chest compression group was significantly lower than that of the manual chest compression group (OR = 0.24, 95% CI 0.14–0.41, Z = 5.12, p < 0.05). The forest plot analysis is illustrated in Fig. 5.

Fig. 5.

Incidence of rib fractures forest plot of Thumper chest compression versus manual chest compression. CPR, cardiopulmonary resuscitation; CI, confidence interval.

3.6 Publication Bias

The ROSC funnel plot shape was inverted and symmetrical, indicating no significant bias (Fig. 6). The symmetry of the survival rate and rib fracture funnel plots could not be evaluated due to limited studies evaluating these outcomes. However, when considering the small sample size in these studies, publication bias could not be ruled out.

Fig. 6.

Funnel plot of the included paper. OR, odds ratio; SE, standard error.

3.7 Sensitivity Analysis

The sensitivity analysis forest plots are shown in Figs. 7,8,9. The sensitivity analysis showed that the difference between before and after each effect index was small. The difference between the relative risk (RR) and OR models was not obvious. The outcomes of the study of None et al. [38] varied when compared with the other studies, in order to exclude bias caused by language differences and it was therefore excluded from the sensitivity analysis. The results of the effect indicators for the ROSC, survival rate, and incidence of rib fractures were 2.65 (95% CI 2.17–3.23, Z = 9.65), 4.24 (95% CI 2.87–6.26, Z = 7.28), and 0.18, (95% CI 0.10–0.35, Z = 5.22), respectively. For all outcome indicators, the results were statistically significant (p < 0.00001), indicating that the findings of the meta-analysis are highly reliable.

Fig. 7.

The ROSC Subgroup analysis forest plot of Thumper device group versus manual CPR group. CPR, cardiopulmonary resuscitation; CI, confidence interval; ROSC, return of spontaneous circulation.

Fig. 8.

The survival rate Subgroup analysis forest plot of Thumper device group versus manual CPR group. CPR, cardiopulmonary resuscitation; CI, confidence interval.

Fig. 9.

The rib fracture rate Subgroup analysis forest plot of Thumper device group versus manual CPR group. CPR, cardiopulmonary resuscitation; CI, confidence interval.

4. Discussion

The application of CPR for CA patients aims to restore spontaneous circulation as soon as possible and to obtain better neurological function after discharge. The 2015 AHA guidelines for CPR emphasize that early high-quality CPR is crucial to improving survival and neurological function in CA patients [6]. Current methods of compression include manual compression and mechanical compression, each of which has advantages and disadvantages. However, current study provide controversial findings. Couper et al. [11] showed that chest compression provided a better 30-day survival rate and short-term survival rate when compared with manual chest compression. In addition, Westfall et al. [21] showed that mechanical chest compression could significantly improve ROSC and load compression distribution compared with manual compression. However, only a few study showed that mechanical chest compression could significantly improve the prognosis of CA patients [11, 21]. In contrast, several study conducted in recent years with more participants concluded that the use of mechanical compression devices did not improve prognosis compared with manual chest compressions [1, 15]. In a study performed by Zhu et al. [14], the ROSC rate, in-hospital survival rate, discharge survival rate, and CPC score in the manual chest compression group were significantly better than those of the mechanical chest compression group. The study conducted by Khan et al. [15] showed that mechanical chest compression performed resulted in a similar 30-day survival rate, discharge survival rate, admission survival rate, ROSC, neurological function recovery, and rib fracture incidence when compared with manual chest compressions. However, although some studies concluded that mechanical compression has no obvious advantage in the prognosis of CA patients, it can be difficult to provide high-quality chest compression in some special circumstances, such as during transportation, operating conditions, and in situations whereby the safety of the rescuers is at risk. Therefore the use of mechanical chest compression is still high.

To our knowledge, there is currently no meta-analysis comparing the survival and prognosis between the Thumper and manual chest compressions.

A total of 16 RCTs were included in this meta-analysis. The results showed that there were differences between manual compression and mechanical compression in terms of the three observed indicators (the ROSC rate, the rate of survival to hospital discharge, and Incidence of rib fractures). There were several methodological differences between studies that could have influenced the findings of our meta-analysis, such as the number of times, duration, and locations of the CPR. Furthermore, mechanical chest compression devices and CPR practices are constantly improving, potentially explaining the worse outcomes noted in earlier studies such as the one by Taylor et al. [38].

The results of this meta-analysis differ from the findings of other published meta-analyses. Liu et al. [1] evaluated the data of six studies with a total of 8501 participants, and found that the use of LUCAS chest compression devices did not improve ROSC (OR = 1; 95% CI: [0.89, 1.13]) and hospital survival (OR = 0.86; 95% CI: [0.65, 1.15]). Bonnes et al. [17] conducted a large-sample review (n = 9157) of data from observational studies. They concluded that the use of mechanical devices could significantly improve short-term outcomes such as ROSC and discharge survival. Still, no significant benefit was observed in the long-term prognosis after discharge.

Three reasons could explain the different findings of our meta-analysis. In previous meta-analyses, chest compression devices with a three-dimensional compression mode were used as opposed to the single-point compression mode used by the Thumper device. The Cochrane quality assessment classified the studies as very low in quality or uncertain. Meta-analyses based on low-quality studies may overestimate or underestimate the effectiveness of treatment [39]. Therefore the current evidence is not sufficient to support the effect of the Thumper device of resuscitation, particularly for the assessment of survival and incidence of rib fractures. Finally, differences in the operating environment could lead to differences in resuscitation outcomes. Spiro et al. [40] and Parnia et al. [41] have shown that mechanical devices are more effective in resuscitation and maintaining patients’ survival than manual chest compressions for CA patients in the hospital environment. This may be related to the ability to provide greater team support in-hospital than out-of-hospital after cardiac arrest. Moreover, mechanical chest compression devices can be deployed earlier in the hospital environment, potentially improving outcomes. However, the meta-analysis of Bonnes et al. [17] has shown that the earlier the mechanical device is deployed during a CA in the out-of-hospital environment, the more effective the CPR is. It is important to note that chest compression pauses related to the deployment of a mechanical chest compression device are rarely reported. Studies have shown that the rescuer’s skills can have an influence on the deployment speed and performance of mechanical chest compression, which could also have an impact on outcomes [11, 42]. In addition, in the hospital environment, manual chest compression is usually difficult to implement because patients are usually positioned on a compressible mattress, which can absorb up to 40% of the compression force, resulting in a lower compression depth than the standard required by the AHA guidelines [43].

The most important strength of this system review and meta analysis is the first analysis of the survival of Thumper mechanical chest compression. This is of clinical relevance with regard to increasing treatment options in hospital, where automated mechanical CPR devices might provide a “bridge” to definitive treatment for designated patient groups under certain conditions and in specific environments. We hope that this systematic review and meta- analysis may contribute to implicate future clinical and scientific issues towards an individualised decision making.

This meta-analysis has some limitations that have to be acknowledged. The overall Cochrane quality evaluation showed that the quality of the RCTs evaluated in the study was often low. Not all RCTs evaluated in this study explained the technique used to randomize the participants into the mechanical and manual compression groups and the blinding method used to evaluate clinical outcomes. Details on the patients lost to follow-up were not always provided. The limited number of studies included in this meta-analysis and the small sample size of the included studies may limit the generalizability of the research findings. Therefore further high-quality RCTs are recommended to confirm the effect of the Thumper device. In addition, this study only evaluated the outcomes of the Thumper chest compression in relation to manual chest compression in CA patients, highlighting the need for further research to evaluate the outcomes of other mechanical chest compression devices. It is also important to note that long-term survival is affected by many factors, such as the severity of CA, patient co-morbidities, and follow-up treatment. Therefore these findings may not reflect the long-term survival outcomes, highlighting the need for further longitudinal studies. Finally, this meta-analysis did not evaluate the clinical impact of other co-founding variables such as etiology and location of the CA procedure (e.g., within a hospital or outside the hospital), and hence more research is required to assess the impact of these variables on clinical outcomes to better guide rescuers on the use of mechanical chest compression in CA patients.

5. Conclusions

In this review, our meta-analysis found an association between improved hospital survival and treatment with a Thumper compression device for in-hospital cardiac arrest. We also found evidence of improved short-term survival and improved physiological outcomes when a mechanical device was used. Nevertheless, it is important to note that the Thumper device may not always be available to the rescuers. This review suggests a potential role for mechanical chest compression devices for in-hospital cardiac arrest, but there is an urgent need for high-quality research, particularly adequately powered randomised trials, to further examine this role.

Author Contributions

Study design—DL, YW; Data collection—NZ, JW; Data analysis—HZ, BX; Study supervision—DL, JW; Manuscript writing: Critical revisions for important intellectual content—HZ, JW. All authors contributed to editorial changes in the manuscript. All authors read and approved the final manuscript. The authors agree to take responsibility for the work.

Ethics Approval and Consent to Participate

Not applicable.

Acknowledgment

We would like to express our gratitude to all those who helped us during the writing of this manuscript. Thanks to all the peer reviewers for their opinions and suggestions. We would like to thank TopEdit (www.topeditsci.com) for the English language editing of this manuscript.

Funding

This research was funded by Hainan Provincial Natural Science Foundation of China (822MS071), Chinese Academy of Medical Sciences Innovation Fund for Medical Sciences (2019-I2M-5-023), Key Laboratory of Emergency and Trauma (Hainan Medical University), Ministry of Education (Grant. KLET-202105), Youth Program of Naval Medical University Basic Medical Fund (2021QN16), Hainan Provincial Science and Technology Major Project (ZDKJ202004, ZDKJ2021038) and 2022 Hainan Province Philosophy and Social Science Planning project (HNSK(QN)22-84).

Conflict of Interest

The authors declare no conflict of interest.

References
[1]
Liu M, Shuai Z, Ai J, Tang K, Liu H, Zheng J, et al. Mechanical chest compression with LUCAS device does not improve clinical outcome in out-of-hospital cardiac arrest patients: A systematic review and meta-analysis. Medicine. 2019; 98: e17550.
[2]
Berdowski J, Berg RA, Tijssen JGP, Koster RW. Global incidences of out-of-hospital cardiac arrest and survival rates: Systematic review of 67 prospective studies. Resuscitation. 2010; 81: 1479–1487.
[3]
Gräsner JT, Herlitz J, Koster RW, Rosell-Ortiz F, Stamatakis L, Bossaert L. Quality management in resuscitation–towards a European cardiac arrest registry (EuReCa). Resuscitation. 2011; 82: 989–994.
[4]
Hunyadi-Antievi S, Proti A, Patrk J, Filipovi-Gri B, Radulovi B. European Resuscitation Council Guidelines for Resuscitation 2015. Lijecnicki vjesnik. 2018; 138: 305–321.
[5]
Zègre-Hemsey JK, Bogle B, Cunningham CJ, Snyder K, Rosamond W. Delivery of Automated External Defibrillators (AED) by Drones: Implications for Emergency Cardiac Care. Current Cardiovascular Risk Reports. 2018; 12: 25.
[6]
Panchal AR, Berg KM, Hirsch KG, Kudenchuk PJ, Del Rios M, Cabañas JG, et al. 2019 American Heart Association Focused Update on Advanced Cardiovascular Life Support: Use of Advanced Airways, Vasopressors, and Extracorporeal Cardiopulmonary Resuscitation During Cardiac Arrest: An Update to the American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2019; 140: e881–e894.
[7]
Casner M, Andersen D, Isaacs SM. The impact of a new CPR assist device on rate of return of spontaneous circulation in out-of-hospital cardiac arrest. Prehospital Emergency Care. 2005; 9: 61–67.
[8]
Ikeno F, Kaneda H, Hongo Y, Sakanoue Y, Nolasco C, Emami S, et al. Augmentation of tissue perfusion by a novel compression device increases neurologically intact survival in a porcine model of prolonged cardiac arrest. Resuscitation. 2006; 68: 109–118.
[9]
Halperin HR, Paradis N, Ornato JP, Zviman M, Lacorte J, Lardo A, et al. Cardiopulmonary resuscitation with a novel chest compression device in a porcine model of cardiac arrest: improved hemodynamics and mechanisms. Journal of the American College of Cardiology. 2004; 44: 2214–2220.
[10]
Magliocca A, Olivari D, De Giorgio D, Zani D, Manfredi M, Boccardo A, et al. LUCAS Versus Manual Chest Compression During Ambulance Transport: A Hemodynamic Study in a Porcine Model of Cardiac Arrest. Journal of the American Heart Association. 2019; 8: e011189.
[11]
Couper K, Yeung J, Nicholson T, Quinn T, Lall R, Perkins GD. Mechanical chest compression devices at in-hospital cardiac arrest: A systematic review and meta-analysis. Resuscitation. 2016; 103: 24–31.
[12]
Gates S, Quinn T, Deakin CD, Blair L, Couper K, Perkins GD. Mechanical chest compression for out of hospital cardiac arrest: Systematic review and meta-analysis. Resuscitation. 2015; 94: 91–97.
[13]
Obermaier M, Zimmermann JB, Popp E, Weigand MA, Weiterer S, Dinse-Lambracht A, et al. Automated mechanical cardiopulmonary resuscitation devices versus manual chest compressions in the treatment of cardiac arrest: protocol of a systematic review and meta-analysis comparing machine to human. BMJ Open. 2021; 11: e042062.
[14]
Zhu N, Chen Q, Jiang Z, Liao F, Kou B, Tang H, et al. A meta-analysis of the resuscitative effects of mechanical and manual chest compression in out-of-hospital cardiac arrest patients. Critical Care. 2019; 23: 100.
[15]
Khan SU, Lone AN, Talluri S, Khan MZ, Khan MU, Kaluski E. Efficacy and safety of mechanical versus manual compression in cardiac arrest - A Bayesian network meta-analysis. Resuscitation. 2018; 130: 182–188.
[16]
Li H, Wu J, Zhang Y, Jing X. GW29-e0549 Meta-analysis of mechanical versus manual chest compressions in cardiopulmonary resuscitation. Journal of the American College of Cardiology. 2018; 72: C243–C244.
[17]
Bonnes JL, Brouwer MA, Navarese EP, Verhaert DVM, Verheugt FWA, Smeets JLRM, et al. Manual Cardiopulmonary Resuscitation Versus CPR Including a Mechanical Chest Compression Device in Out-of-Hospital Cardiac Arrest: A Comprehensive Meta-analysis From Randomized and Observational Studies. Annals of Emergency Medicine. 2016; 67: 349–360.e3.
[18]
Li H, Wang D, Yu Y, Zhao X, Jing X. Mechanical versus manual chest compressions for cardiac arrest: a systematic review and meta-analysis. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine. 2016; 24: 10.
[19]
Tang L, Gu W, Wang F. Mechanical versus manual chest compressions for out-of-hospital cardiac arrest: a meta-analysis of randomized controlled trials. Scientific Reports. 2015; 5: 15635.
[20]
Brooks SC, Hassan N, Bigham BL, Morrison LJ. Mechanical versus manual chest compressions for cardiac arrest. The Cochrane Database of Systematic Reviews. 2014; CD007260.
[21]
Westfall M, Krantz S, Mullin C, Kaufman C. Mechanical versus manual chest compressions in out-of-hospital cardiac arrest: a meta-analysis. Critical Care Medicine. 2013; 41: 1782–1789.
[22]
Higgins JP, Thompson SG, Deeks JJ, Altman DG. Measuring inconsistency in meta-analyses. British Medical Journal. 2003; 327: 557–560.
[23]
Lingfang M. Application of Thumper cardiopulmonary resuscitation device in emergency patients with cardiac arrest. Medical Journal of Chinese People’s Health. 2019; 31: 19–20.
[24]
He NN, Wu DH, Li K, Hu BJ. The effectiveness of Thumper 1007 on cardiopulmonary resuscitation and the key indicators. Modern Preventive Medicine. 2016; 43: 3049–3052.
[25]
Gong L, Sun BL, Zhao YL, Fan Y, Wu TT, Chen YQ. Therapeutic effect of Thumper cardiopulmorary resuscitation system in the emergency treatment with sudden cardiac arrest. China of Heart Journal. 2016; 28: 194–196.
[26]
Ding HB. Clinical application of Thumper machine for cardiopulmonary resuscitation. World’s Latest Medical Information Abstracts. 2014; 27:57-59.
[27]
Zheng H, Ye FX, Ye N, Fu XL, Zhang L. Application effect of Saab cardiopulmonary resuscitation machine on rescue of patients with cardiac arrest. Chinese Youjiang Medical Journal. 2019; 47: 584–586. (In Chinese)
[28]
Zhang CY, Wang Y, Liu H, Hao A, Xun J, Meng Q. Comparative study for effects of bare-handed CPR with Thumper cardiopulmonary resuscitator CPR in emergency department based on “the Utstein style”. Chinese Critical Care Medicine. 2017; 937–939. (In Chinese)
[29]
You Y MY. Nursing Research on Application of Thumper 1007 device in Cardiopulmonary Resuscitation. Journal of Xuzhou Medical University. 2017; 37: 692–693.
[30]
Qiaoli D. Analysis of clinical use of Thumper CPR device in emergency. Cardiovascular Disease Journal of Integrated Traditional. 2016; 4: 83–85.
[31]
Liu H, Zheng H, Zheng W, Fu XL, Li ZX. Effect analysis of Thumper cardiopulmonary resuscitation device during the cardiac arrest. Acta Medicinae Sinica. 2015; 28: 5–8.
[32]
Guo S, Gong DB, Lu XG, Bai LZ, Fan ZW, Wang YL. Analysis of Thumper Cardiopulmonary Resuscitation Machine in the Emergency Rescue. China Medical Device Information. 2014;20: 3–5+27.
[33]
Yang J, Lixia H, Zhang N. Observation on the effect of saber cardiopulmonary resuscitation machine in clinical cardiopulmonary resuscitation. China of Baotou Medicine. 2013; 37: 98–99.
[34]
Liu J, Duan M, Wei S, Feng C. Evaluation of effect of two CPR means by end - tidal carbon dioxide partial pressure. Journal of Clinical and Experimental Medicine. 2013; 12: 103–104.
[35]
Hu PB, Ling XU, Qiu JQ, Yi L. Efficacy analysis of cardiopulmonary resuscitation with Thumper-cardiopulmonary resuscitation device. China Journal Lung Disease (Electronic Flition). 2012; 5: 245–247.
[36]
Huang Q. Effect observation of cardiopulmonary resuscitation by Thumper CPR machine. Chinese and Foreign Medical Research. 2011; 9: 41–42.
[37]
Lu XG, Kang X, Gong DB, Fan ZW, Bai LZ, Dong Y, et al. The clinical efficacy of Thumper modal 1007 cardiopulmonary resuscitation: a prospective randomized control trial. Chinese Critical Care Medicine. 2010; 22: 496–497.
[38]
Taylor GJ, Rubin R, Tucker M, Greene HL, Rudikoff MT, Weisfeldt ML. External cardiac compression. A randomized comparison of mechanical and manual techniques. The Journal of the American Medical Association. 1978; 240: 644–646.
[39]
McFadden P, Reynolds JC, Madder RD, Brown M. Diagnostic test accuracy of the initial electrocardiogram after resuscitation from cardiac arrest to indicate invasive coronary angiographic findings and attempted revascularization: A systematic review and meta-analysis. Resuscitation. 2021; 160: 20–36.
[40]
Spiro JR, White S, Quinn N, Gubran CJ, Ludman PF, Townend JN, et al. Automated cardiopulmonary resuscitation using a load-distributing band external cardiac support device for in-hospital cardiac arrest: a single centre experience of AutoPulse-CPR. International Journal of Cardiology. 2015; 180: 7–14.
[41]
Parnia S, Nasir A, Ahn A, Malik H, Yang J, Zhu J, et al. A feasibility study of cerebral oximetry during in-hospital mechanical and manual cardiopulmonary resuscitation*. Critical Care Medicine. 2014; 42: 930–933.
[42]
Baloglu Kaya F, Acar N, Ozakin E, Canakci ME, Kuas C, Bilgin M. Comparison of manual and mechanical chest compression techniques using cerebral oximetry in witnessed cardiac arrests at the emergency department: A prospective, randomized clinical study. The American Journal of Emergency Medicine. 2021; 41: 163–169.
[43]
Perkins GD, Kocierz L, Smith SCL, McCulloch RA, Davies RP. Compression feedback devices over estimate chest compression depth when performed on a bed. Resuscitation. 2009; 80: 79–82.

Publisher’s Note: IMR Press stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.