Reviews in Cardiovascular Medicine (RCM) is published by IMR Press from Volume 19 Issue 1 (2018). Previous articles were published by another publisher in Open Access under a CC-BY (or CC-BY-NC-ND) licence, and they are hosted by IMR Press on imrpress.com as a courtesy and upon agreement with MedReviews, LLC.

Open Access Review
Implantable Cardioverter Defibrillator Shocks: A Troubleshooting Guide
Show Less
1 Division of Cardiology, Cedars-Sinai Medical Center, Los Angeles, CA
Rev. Cardiovasc. Med. 2001, 2(2), 61–72;
Published: 30 June 2001
Abstract
Implantable cardioverter defibrillators deliver shocks in response to electrical signals that satisfy programmed criteria for detection of VT or VF. The first step in diagnosis of inappropriate shocks in patients with ICDs is to determine if the shock was delivered in response to a true tachyarrhythmia by inspecting data stored in the ICD. Shocks occur in the absence of tachyarrhythmias because nonarrhythmic physiologic or nonphysiologic signals are oversensed by the ICD and detected as arrhythmias. Diagnosis and causes of oversensing are reviewed. The second step in diagnosis is to determine if the tachyarrhythmia stored in the VT/VF episode log is VT/VF or SVT by analyzing stored electrograms. Frequent or repetitive shocks constitute an electrophysiologic emergency. The approach to this problem is reviewed.
Keywords
Implantable cardioverter defibrillator
Ventricular tachycardia
Inappropriate shocks