Academic Editor: Sergio Bagnato
Background: Deviation of electrode array from the midline of spinal
cords affects the therapeutic outcomes of C2-4 cervical spinal
cord stimulation (SCS) in patients with disorders of consciousness (DOC).
This study proposed the implementation of a novel C2-3 dural
exposure procedure and investigated its efficacy compared to conventional
surgery. Methods: Surgical and postoperative imaging data from 69
patients with DOC who underwent SCS in the lateral decubitus position were
retrospectively assessed. The C2-3 dural exposure procedure was performed in 16
patients while the rest underwent conventional surgery. The incidence of
electrode deviation was compared, and factors associated with the deviation were
investigated. Results: The rate of complete midline coverage by the
electrodes in the C2-3 dural exposure group was significantly higher than the
conventional group (93.8% vs. 54.7%, p = 0.004). Exposure of the dura
between C2-3 was a significant favorable factor for complete midline coverage by
the electrode array (odds ratio [OR]: 0.091; 95% confidence interval [CI]:
0.011–0.757; p = 0.027). Electrode positioned
High cervical spinal cord stimulation (SCS) is used in the management of disorders of consciousness (DOC) [1, 2, 3], especially in patients in a minimally conscious state. It has been shown to have an important role in modulating the complexity of neural activities and promoting functional recovery [4, 5]. Its target stimulus area is the dorsal columns of the C2-4 vertebrae. Traditionally, C5 laminectomy has been performed in the prone position to advance the electrode array craniad to the target zone, where a pulse generator is implanted in the lateral abdominal region [6]. However, one limitation of this procedure we observed in clinics is the challenge of ensuring that the electrode has completely covered the midline of the spine.
When high cervical stimulation is used in pain management, its most common indication, the clinical determination of the electrode position is accomplished under local anesthesia. However, intraoperative sensory tests cannot be performed in patients with DOC as they cannot effectively respond to stimuli. Thus, to increase the effectiveness of therapy, accurate implantation of electrodes at predetermined locations is a prerequisite. However, literature on the technical aspects of cervical SCS implantation is scarce, and few studies have documented the accuracy of DOC-SCS implantation.
In this study, we report our findings on an improved paddle-type C2-4 spinal cord stimulator implantation technique in patients with DOC operated in the lateral decubitus position. A C2-3 dural exposure procedure was adopted to overcome the limitation of previous surgical methods, such as the inability to ensure complete midline coverage. Then, we compared the rate of complete midline coverage by the electrode array in patients with DOC who underwent conventional surgery with those who received the proposed C2-3 dural exposure procedure and further investigated the determinants of complete midline coverage by the electrode array.
In this retrospective single-center study, the data of adult (
The surgeries were performed under general anesthesia and intubation by the same
team of surgeons with more than six years of experience. For patients with cuffed
tracheostomy cannula, the cannula was fixed on the neck with medical polyurethane
and was used for interoperative care and ventilation. The pulse generator was
placed in front of the pectoralis major fascia under the left or right clavicle.
Sides with a ventriculoperitoneal shunt were avoided. The patients were then
placed in the left or right decubitus position during the
operation, with the pulse generator opposite to the side that was in contact with
the table. The standard 90° lateral decubitus
positioning was adopted to facilitate manipulation of both the nape of the neck
and the chest wall, with the neck flexed without any axial rotation. For patients
with cervical extensor spasms, their head was fixed with a Mayfield head holder.
Under fluoroscopy, the C5 spinous process was marked using methylene blue. After
the neck and chest regions were disinfected, a midline incision was made from the
C3 to C5 spinous process. C5 laminotomy was performed as previously described
[6]. In patients who underwent conventional surgery (n = 53), a 16-electrode
array (Specify 2
The patients’ medical records, imaging reports and surgical logs were obtained from the hospital’s electronic records system. Data on age, sex, laterality of decubitus position, Glasgow Coma Scale score, CRS-R score, operative time and estimated blood loss were collected and analyzed. Operative time was calculated as the time from entering to leaving the operating room. The estimated blood loss was calculated by weighting the intraoperative surgical gauzes and measuring the volume of the suction fluid. Computed Tomography (CT) was performed using a 64-slice CT scanner (Discovery CT750 HD; GE Healthcare, Chicago, IL, USA) within 1 week after surgery. Three-dimensional (3D) CT reconstruction of the cervical spine was also performed. The sagittal diameter of the spinal canal at the lower edge of the C2 vertebra was measured on CT. Full coverage of the midline by the electrode array was defined as the area between the lateral sides of the two contacts at each level covering the dorsal midline of the spine, which was the primary study outcome. It was determined based on the axial scan of the cephalic-to-caudal contact pairs. The deviation direction was also recorded (Fig. 1). The height of the electrode was obtained by measuring the height of most cephalic contacts above the lower edge of C2.
Representative images of the surgical positioning and postoperative computed tomography (CT). (A) A patient in the right decubitus position, with incision markings at the C2-5 spinous processes. (B–D) Representative images of the electrode array in the coronal (B), horizontal (C) (at the most cephalic contact level), and sagittal (D) CT views.
Electromyography was used during device programming approximately two weeks
after surgery in some studied patients (n = 45). Contact of interest (COI) was
selected as a contact or contact pair on the midline at the cephalic level. The
stimulus paradigm was set to a frequency of 70 Hz and a pulse width of 210
Prognosis of consciousness was assessed in all the patients. The CRS-R score was recorded 1 month after surgery during hospitalization or doctor’s home visits. The Extended Glasgow Outcome Scale (GOSE) score was recorded during follow-up via telephone or re-examination by an experienced neurologist 1 year after surgery.
Statistical analyses were performed using the SPSS software (version 24.0; IBM,
Armonk, NY, USA). The data were expressed as mean (
The study flowchart of patient data collection is shown in Fig. 2. Initially,
the records of 98 patients diagnosed with DOC were screened. Of them, 76 patients
underwent SCS. Four cases were excluded because other types of electrode arrays
were used, or the patients were placed in a prone position. Two cases were
excluded due to the lack of postoperative imaging. One case was excluded because
the age did not match the study criteria. Finally, 69 cases were eligible for
this study, of whom 53 underwent conventional surgery, and 16 received the C2-3
dural exposure procedure. Their mean age was 43.5
The study flowchart. Abbreviations: DOC, disorders of consciousness; SCS, spinal cord stimulation.
Characteristics | Conventional surgery (n = 53) | C2-3 dural exposure (n = 16) | t/ |
p | |
---|---|---|---|---|---|
Male sex, n (%) | 41 (77.4) | 11 (68.9) | 0.136 | 0.712 | |
Age, years | |||||
Mean |
42.6 |
46.7 |
1.086 | 0.282 | |
Range | 18–65 | 23–69 | |||
Duration of DOC, months | |||||
Median (range) | 4.0 (3–41) | 5.5 (3–156) | |||
DOC types, n (%) | |||||
Vegetative state | 41 (77.4) | 12 (75.0) | |||
Minimally conscious state | 12 (22.6) | 4 (25.0) | |||
Etiologies, n (%) | |||||
Traumatic | 15 (28.3) | 6 (38.5) | |||
Anoxic | 16 (30.2) | 0 (0) | |||
Vascular | 22 (42.5) | 10 (62.5) | |||
Abbreviations: DOC, disorders of consciousness. |
Electrodes and pulse generators were implanted in one stage in all patients of
both treatment groups. No adjustments of electrodes were performed for serious
adverse reactions to stimulation or other causes. The mean operative time was
170.9
At the time of device programming, two weeks after surgery, 45 patients underwent EMG testing. CT scans showed that in 30 patients, the contacts at all levels fully covered the midline, 1 (3.3%) patient had EMG lateralization, while 15 patients had cephalic contacts deviated from the midline, of which all were lateralization. The average voltage at which laterality was detected was 4.5 V.
At 1 month after surgery, 23 (27.5%) patients had improved CRS-R scores, including 13 who underwent conventional surgery and 6 who underwent the C2-3 dural exposure procedure. One year after surgery, the GOSE scores were 4 points in 1 patient, 3 points in 14 patients, 2 points in 52 patients, and 1 point in 2 patients. No significant difference in scores was observed between the two treatment methods (Table 2).
Characteristics | Conventional surgery (n = 53) | C2-3 dural exposure (n = 16) | t/ |
p | |
---|---|---|---|---|---|
Spinal canal sagittal diameter, mm | |||||
Mean |
15.1 |
14.9 |
0.539 | 0.592 | |
Range | 11.1–18.8 | 12.7–19.3 | |||
Electrode height, mm | |||||
Median (range) | 5.9 (–11.2 to 18.2) | 2.3 (–6.5 to 10.8) | 0.94 | 0.351 | |
Midline complete coverage (%) | 29 (54.7) | 15 (93.8) | 8.104 | 0.004* | |
Operating room time, min | |||||
Mean |
170.4 |
172.8 |
0.970 | 0.336 | |
Range | 160–190 | 155–195 | |||
Estimated blood loss, mL | |||||
Median (range) | 50 (10–200) | 50 (50–150) | 1.113 | 0.270 | |
Surgical site infection, n | 0 | 0 | |||
1-month CRS-R improvement (%) | 13 (24.5) | 6 (37.5) | 0.488 | 0.485 | |
1-year GOSE score | |||||
1 (%) | 1 (1.9) | 1 (6.2) | |||
2 (%) | 43 (81.1) | 9 (56.2) | |||
3 (%) | 8 (15.1) | 6 (37.5) | |||
4 (%) | 1 (1.9) | 0 (0) | |||
5–8 (%) | 0 (0) | 0 (0) | |||
GOSE score |
9 (17.0) | 6 (37.5) | 1.955 | 0.162 | |
Abbreviations: *, statistical significance (p Nine patients had a second cervical CT scan between 6 months and 2 years postoperatively, with 3 patients underwent conventional surgery and 6 patients underwent surgery with C2-3 dural exposure. No displacement, migration or breakage of the electrodes was found. |
In this study, we observed a trend towards a greater chance of deflection at
higher electrode positioning. In addition, electrodes placed in vertebrae of
smaller diameter seemed to be displaced due to potential resistance exerted on
them. Therefore, electrode positioning, spinal diameter and surgical method were
selected as regression variables. Logistic regression analysis revealed that
exposure of dura between C2-3 was a significant favorable factor for complete
midline coverage by the electrode array (odds ratio [OR]: 0.091; 95% confidence
interval [CI]: 0.011–0.757; p = 0.027) (Table 3). Meanwhile, electrodes
positioned
Variables | OR (95% CI) | p |
---|---|---|
Electrode position ( |
1.126 (1.016, 1.248) | 0.023* |
C2-3 dural exposure vs. no exposure | 0.091 (0.011, 0.757) | 0.027* |
Spinal canal sagittal diameter ( |
1.058 (0.740, 1.514) | 0.756 |
*Indicates statistical significance (p |
In this study, we proposed a C2-3 dural exposure procedure and compared its efficacy with conventional surgery. Our findings showed that although the caudal side of the electrode could be fixed in the midline, the electrode tip would still be skewed to one side in some patients, as viewed on early postoperative CT scans. These contacts that deviated from the midline would often produce a jolting stimulation corresponding to dorsal root involvement, resulting in patients’ discomfort and limiting the selection of program control parameters [8]. Specifically, according to the overlapping percentage of the contacts, complete midline coverage could be divided into two cases: (1) the contacts pair is either symmetrically distributed on both sides of the midline or (2) only one contact in a pair covers the midline (Fig. 3). Symmetrical stimulation is easier to achieve with the former, while the latter requires unilateral contact stimulation to ensure bilateral effects.
Diagrammatic coronal view depicting the electrode array placement. (A) The electrode array enters the epidural space through the laminectomy level, while the cephalic contacts deviate from the midline. (B) Dural exposure between C2-3 ensures full coverage of the midline by the electrode array.
The proposed C2-3 dural exposure procedure was associated with a significant reduction in cephalic contact deviation (Fig. 3) and solved difficulties in programming. Compared with the traditional method of C5 laminotomy and intraoperative adjuncts, including fluoroscopy, the C2-3 dural exposure procedure has the following advantages: (1) the electrode array can be implanted under direct vision, thereby avoiding parallax and visual alignment errors during photography; (2) it is easier to separate the epidural space at the C2-3 level to adjust the electrode array if the surgeon is not satisfied with the initial placement site. Such adjustment is difficult when using the traditional method because the first attempt often forms a misguided channel within the epidural ligaments, and; (3) potential dural laceration could be avoided by dissecting the dorsal meningovertebral ligaments under direct vision.
A correlation between electrode height and complete midline coverage by the electrode array was observed in this study. The height of the electrode was calculated from the lower edge of C2 because a force of lateral slippage would often be experienced as the electrode array passes below the C2 spinous process. In the prone position, the same obstacle can be encountered. This is not attributable to pathological spinal stenosis but to the complexity of high cervical SCS [9], which can often lead to complications and revisions of electrode placements. As upper cervical segments contribute substantially to the rotational movements of the neck, the epidural space contains intricate connections that merge and blend together [10]. Meningovertebral ligaments connecting the posterior dural sac to the ligamentum flavum or the lamina are strong and maintain the integrity of this region. Cadaver studies showed that the ligaments were high in collagenous tissues [10] and were thickest at the C1 and C2 vertebrae [11]. Our observation concurs with these findings. The issues related to epidural resistance at the C2 level could be largely resolved by implementing the C2-3 dural exposure procedure.
Our findings also showed that surgical exposure adjustments did not significantly prolong the operation time nor increase the amount of bleeding. Also, no surgical site infection was observed between the 2 treatment groups. Epidural resistance was encountered around C2, but this was physiologic rather than scarring. With the C2-3 dural exposure technique, blind exploration of the epidural can be avoided, thereby increasing safety. Despite extending the wound cephalically, there were no wound healing problems, and no significant difference was found in postoperative incision pain. Compared to techniques that require additional laminectomies in the presence of epidural resistance or adhesions [12], the proposed C2-3 dural exposure procedure was associated with less damage to bony structures and retained most muscle attachment points, especially at the C2 spinous process. This can help maintain neck stability and reduce the impact of surgery on muscle tone. Interestingly, considering that a significant proportion of patients with DOC often suffer from spasm problems, in this study, we observed a short-term relief of neck or upper limb spasms in both surgery groups. Only one patient who underwent conventional surgery developed a prolonged weakness in neck extension. The symptoms improved a lot during the subsequent six months of rehabilitation therapy.
As the patients spent more time in bed, stimulator displacement in the upright position was not of concern in this study. Hence, we preferred placing the pulse generator into the chest wall rather than the flank or upper buttock area to reduce the length of the extension wire and potential injuries during passive turning over movements.
The lateral decubitus position was adopted in this study. This position allows a one-time skin disinfection followed by electrode implantation in the neck and pulse generator implantation in the chest wall, which can ultimately reduce the operation time and risk of infections. Another reason for choosing the lateral decubitus was to facilitate the anesthesiologist to check the tracheotomy cannula, which is not readily visible in the prone position. The neck position and operator’s perspective could make the electrode deviate to the decubitus side. We, therefore, emphasize that the incision should be perpendicular to the dermatoglyph of the neck to avoid lateral flexion and torsion of the neck, and the bed should be adjusted before the operation to ensure that it is at the surgeon’s eye level for incision. Thus, most deviation from the midline could be corrected with the proposed C2-3 dural exposure.
Compared to the single-column paddle, a multi-column paddle increases the
coverage area and the flexibility of program control. In our previous attempts,
we noted that the Medtronic Specify 2
In addition to direct observation, there are two methods for evaluating the location of the electrode array during general anesthesia. One is intraoperative imaging. Fluoroscopy is widely used to locate the spinal segment and plan the skin incision. It also allows for some minimal access to spinal technologies that can be used when the patient is conscious [12, 14]. However, this requires an experienced surgeon to accurately determine intraoperative midline implants because parallax and visual alignment errors during photography cannot truly reflect the spatial relationships between the electrodes, and the spinal canal of patients with brain damage frequently has cervical lateral flexion and torsion [15]. Therefore, intraoperative CT is the preferred imaging modality. It can also accurately measure the implantation location of the electrodes.
For patients with a deviation of the spinal cord from the midline of the canal, a more desirable method of verification would be neurophysiological monitoring. This includes electromyography and somatosensory-evoked potential collision studies (SSEP-CS), which have recently been used in patients with chronic pain [16, 17, 18]. Electromyography is more sensitive and helps guide device programming [16, 17, 18]. SSEP-CS is safe for the intraoperative monitoring of midline placement of C1-2 paddle leads [19]. The common unilateral brain insufficiency in the patients of this study cohort required caution if SSEP-CS was to be used as an indicator of laterality. Thus, electromyography was preferred for intraoperative monitoring and device programming. C2-3 dural exposure facilitated the adjustment of electrodes according to neurophysiological monitoring. Two weeks after surgery, EMG was used to assess the efficacy of midline coverage, as determined by CT. The electrophysiological results matched with the imaging results, showing no obvious electrode displacement during this time interval.
GOSE was used to assess the long-term recovery of consciousness in this study. Under this scoring system, a score greater than 3 refers to full recovery of consciousness, while a score greater than 4 refers to partial recovery of self-care. Although there was a significant difference in midline coverage between the 2 treatment groups, no significant difference in conscious outcome was observed. There are many prognostic factors for consciousness disorder, including etiology, preoperative consciousness level and others. Long-term prognosis is closely related to the nursing level after discharge. Larger sample sizes may yield more reliable results. Thus, there are ample reasons to necessitate improved midline coverage based on current outcomes. One limitation of this study is that electromyography was performed on only 45 patients at the device programming stage. Therefore, we could not demonstrate the advantages of the C2-3 exposure procedure in terms of electrophysiological monitoring. Another limitation is that the lack of a sufficient number of long-term CT reviews of the electrodes to understand the incidence of long-term electrode displacement, which warrants further research in the future.
This study showed that the proposed C2-4 spinal cord stimulator implantation in patients with DOC under general anesthesia could achieve high midline coverage by: (1) using the lateral decubitus position, (2) exposing the dura between C2-3 to ensure that the electrode array proceeds upwards along the midline, and (3) ensuring that the electrode tip is not too far above the C2 level. These enabled the proposed C2-3 exposure procedure to have lower rates of midline deviation. We hypothesize that this could contribute to easy device programming and reliable efficacy evaluation.
YD, JH and JZ contributed to the study conception and design. YD, XX, YY, RH, JH, JZ collected the data and performed the data analysis. All authors contributed to the interpretation of the data and the completion of figures and tables. All authors contributed to the drafting of the article and final approval of the submitted version.
The study protocol was approved by the Ethics Committee of the Seventh Medical Center of the PLA General Hospital (approval number: 2018-58). The data of all patients were anonymized, and no additional patient consent was required due to the retrospective nature of the study.
Not applicable.
This research received no external funding.
The authors declare no conflict of interest.